T1D glossary › Time in range

Time in range (TIR)

Time in range is the share of CGM values, and of the day, that falls between 70 and 180 mg/dL (3.9 to 10.0 mmol/L). The 2019 international consensus made it one of the core CGM metrics, next to time below range and time above range. It is read over 14 days, with the sensor active at least 70% of that time. (International consensus on time in range, 2019)

Also called time in target range.

In Glide

Target

More than 70% of the time at 70–180 mg/dL (3.9–10.0 mmol/L) for most adults with type 1 or type 2 diabetes; more than 50% for older or high-risk adults. (ADA Standards of Care 2026, section 6; International consensus on time in range, 2019)

Formula

TIR (%) = CGM values from 70 to 180 mg/dL ÷ all CGM values × 100

Worked example

A CGM that records every 5 minutes stores about 4,032 values in 14 days. If 2,903 of them sit between 70 and 180 mg/dL, TIR is 72%, or about 17 hours 17 minutes a day. The consensus paper links each 10% gain in TIR to an A1C about 0.5% lower.

The worked example is our own arithmetic.

Time in range to A1c → The A1c that a time in range predicts, with its 95% range.